Calibration

Thermo Scientific
#1 EU Min Value – This field displays the minimum temperature value, in
engineering units, that was applied by the user to the RTD input during
calibration.
#2 mA Mid Value – This field displays the midpoint temperature value, in
engineering units, that was applied by the user to the RTD input during
calibration.
#3 mA Max Value – This field displays the maximum temperature value, in
engineering units, that was applied by the user to the RTD input during
calibration.
#1 Raw Min Value – This field displays the minimum temperature value read
by the system during calibration.
#2 Raw Mid Value – This field displays the midpoint temperature value read
by the system during calibration.
#3 Raw Max Value – This field displays the maximum temperature value read
by the system during calibration.
Calibration Points – This field displays the number of points the current
information used to calibrate the current input.
Status – This field displays Live or Manual, depending on the operating mode
selected, as well as low and high alarm conditions.
RTD inputs can be calibrated using two or three points. The calibration process for
the RTD inputs should be completed by selecting Calibration from the Functions
menu or by clicking the fourth icon button, the wrench. Selecting either of these
options will guide the user through the calibration process step-by-step. For further
information, see Calibration.
Operation
Detector Screens
4-37
LevelPRO User Manual