Download Print this page

Gauge Calibration - Thermo Scientific LevelPRO User Manual

Advertisement

Startup & the Setup Wizard
The Setup Wizard

Gauge Calibration

-
3
14
LevelPRO User Manual
2. Once the standardization parameters have been completed, click the Next button
to finish standardization.
3. Save the file, if desired, and click the Next button once more to reach the Wizard
Type Selection Screen.
1. Select Gauge Calibration and click the Next button to advance to the following
screen.
Figure 3-13. Gauge Calibration
This screen is used to specify the type of calibration to be performed and the
calibration sample time, in seconds.
CAL Method – The LevelPRO detector performs calibration using a
breakpoint table.
CAL Sample Time (sec) – Like the standardization time, the calibration time
is a period, measured in seconds, during which the instrument will average the
detector counts. Again, the CAL sample time should be based on process
conditions. Typical calibration sample times run between 300 and 900
seconds.
2. Click the Next button to move to the Remaining Time Screen.
Thermo Scientific

Advertisement

loading

This manual is also suitable for:

Levelpro-t