Startup & the Setup Wizard
User Modes & Passwords
User Modes &
Passwords
-
3
4
LevelPRO User Manual
Password protection limits access to the detector's setup parameters based on the access
level rights granted through the four user modes below.
Display List –This is normal mode of operation. The local display will scroll
through a list of measurement data with no ability to view or modify the database
via Remote Backplane (RPB) display/keypad. Remote read access to the database is
available however no writes to the database are allowed.
Operator Mode – The local display on the RBP will be running the user menu
interface, allowing a user to view the configuration of the detector. Remote read
access to the database is available, however no writes to are allowed.
Technician Mode – The local user can configure a small sub-set of the database
within the detector via the front panel menu system on the transmitter. Remote
write access is also available to a small subset of the detector database so that
calibration and standardization can be performed.
Engineer Mode – Local and remote users have full read/write access to the
detector's database.
The password configuration is accessible from the following interface ports:
Com A
Com B
Ethernet
USB
Local Display/Keypad
Fieldbus Interface
Only one port may have access to Engineer mode at any time. The exception to this is
the Fieldbus port, which always has Engineering access, regardless of the operational
modes of the other ports.
The default operation mode is Display List. If no activity is detected on the port for
five minutes, the unit will default back to Display List. See
additional information.
The following values are available for diagnostics purposes, to be used in other
processes and for display purposes.
Com A Password Mode
Com B Password Mode
Ethernet Password Mode
USB Password Mode
Smart Password Mode
Remote Display Password Mode
Current Password Mode
Engineering Mode Count
Operator Mode Count
Technician Mode Count
Invalid Password Count
System Timeout