Figure 4-8. Mode/Fault Alarm Setup, Application Tabs
The following alarms are available for configuration on the Application tabs of the
Mode/Fault Alarm tab.
1. Standardization Mode Active Alarm
a. The alarm is triggered when the level process is in standardization mode.
b. The alarm clears when
i.
Standardization is aborted.
ii. Standardization is accepted.
iii. Standardization is rejected
2. Calibration Mode Active Alarm
a. The alarm is triggered when the level process is in calibration mode.
b. The alarm clears when
i.
Calibration is aborted.
ii. Calibration is accepted.
iii. Calibration is rejected
3. X-Ray Mode Engaged Alarm
a. The alarm is triggered when X-ray engaged is enabled.
b. The alarm clears when X-ray engaged becomes disabled.
4. IBP Communication Failed Alarm
a. The alarm is triggered when the detector is present and an IBP
communication error occurs.
b. The alarm clears when IBP communication errors are cleared.
Operation
Mode/Fault Alarm Setup
4-13
LevelPRO User Manual