Calibration

Thermo Scientific
i.
This selection is not applicable to Level applications.
v. Enable totalizer 4
i.
This selection is not applicable to Level applications.
w. Zero totalizer 4
i.
This selection is not applicable to Level applications.
5. Enter a set point value, in the same units as the selected measurement, in the Set
Point textbox.
6. Enter a clear point value, in the same units as the selected measurement, in the
Clear Point textbox.
7. The following fields display values that are available to be used in other processes.
These fields are for display purposes and are not configurable by the user on this
screen.
Unit – Displays the units of the selected measurement.
Alarm Status – Displays Set or Clear, based on whether the alarm is active or
not.
The LevelPRO detectors have the ability to calibrate all physical inputs and outputs.
To begin input/output calibration, select Calibration from the Functions dropdown
menu at the top of the screen or click the fourth icon button, the wrench.
To access the information on Input/Output Calibration using only the remote
transmitter keypad, see
Figure
A-10,
Figure A-11
and
Figure
Operation
Application Screens
A-12.
LevelPRO User Manual