Operation
Mode/Fault Alarm Setup
-
4
10
LevelPRO User Manual
4. System Fault Alarm
a. The alarm is triggered when any of the following actions occur.
i.
Diagnostic error:
ii. An Input Scan Error occurs and the detector has a problem reading any of
the following devices:
Analog inputs
Main Board Temperature
Detector Analog input
Detector RTD input
b. The alarm clears when the error condition is cleared.
5. Current Output A at Minimum Alarm
a. The alarm is triggered when the value of Current Output A is below the
minimum set value.
b. The alarm is cleared when the value of Current Output A is above the
minimum set value.
6. Current Output A at Maximum Alarm
a. The alarm is triggered when the value of Current Output A is above the
maximum set value.
b. The alarm is cleared when the value of Current Output A is below the
maximum set value.
7. Current Output B at Minimum Alarm
a. The alarm is triggered when the value of Current Output B is below the
minimum set value.
b. The alarm is cleared when the value of Current Output B is above the
minimum set value.
8. Current Output B at Maximum Alarm
a. The alarm is triggered when the value of Current Output B is above the
maximum set value.
b. The alarm is cleared when the value of Current Output B is below the
maximum set value.
9. Current Output C at Minimum Alarm
a. The alarm is triggered when the value of Current Output C is below the
minimum set value.
b. The alarm is cleared when the value of Current Output C is above the
minimum set value.