Calibrate Probe (Cycle971) - Siemens SINUMERIK 840D sl Programming Manual

Measuring cycles
Hide thumbs Also See for SINUMERIK 840D sl:
Table of Contents

Advertisement

Note
Workpiece-related or machine-related measurement requires an appropriately calibrated
tool probe, see Chapter Calibrate probe (CYCLE971) (Page 281).
Compensation strategy
The tool measuring cycle is intended for various applications:
● First measurement of a tool (general setting data SD54762
$SNS_MEA_FUNCTION_MASK_TOOL[Bit9]):
The tool offset values in geometry and wear are replaced.
The offset is applied in the geometry component of the length or the radius.
The wear component is deleted.
● Post measuring a tool (general setting data SD 54762
$SNS_MEA_FUNCTION_MASK_TOOL[Bit9]):
The resulting difference is calculated into the wear component (links or radius) of the tool.
Empirical values may optionally be included. A mean value is not calculated.
3.6.2

Calibrate probe (CYCLE971)

Function
This measuring version calibrates a tool probe machine-related or workpiece-related.
Values are corrected without empirical and mean values.
Measuring cycles
Programming Manual, 12/2017, 6FC5398-4BP40-6BA1
Measuring variants
3.6 Measure tool (milling)
281