During the calibration process, the center point of the calibration ring (corresponding to the
calibration method) and its distance to the starting position are determined.
The calibration data / trigger values are significantly influenced in the result by the following
variables:
● Physical probe ball radius
● Design of the probe
● Measuring velocity
● Calibration ring with corresponding accuracy
● Correct mounting of the calibrating ring
Figure 3-8
Preconditions
The following requirements must be fulfilled for calibration in the ring:
● The probe must be active as tool.
● Probe type:
– 3D multi probe (type 710)
– Mono probe (type 712)
– Star-type probe (type 714)
● Note: The arms of a star-type probe (type 714) must be positioned at 90 degree angle to
each other.
● The exact diameter of the calibration ring is known.
Measuring cycles
Programming Manual, 12/2017, 6FC5398-4BP40-6BA1
Calibrate: Radius in ring (CYCLE976)
Note
In connection with mono probes, star-type probes and the calibration method "Start point
in ring center "NO"", an SPOS-capable spindle is required.
Measuring variants
3.3 Measure workpiece (milling)
113