Operation - Calibration
The following screen appears, which displays the progress of the
injection process on the progress bar.
7. After the process is completed, the Calibration . Stream n . Wizard
4/6 screen appears, on which the actual measured volume must be
entered.
8. On the Calibration . Stream n . Wizard 4/6 screen, enter the actual
measured volume (calibrated vessel) and then select <Confirm>.
With the actual value and the value of the MSC-L measured, a new
meter factor is calculated.
The Calibration . Stream n . Wizard 5/6 screen appears, which
displays the new meter factor.
Fusion4 MSC-L
Installation & Operation Manual
5 - 275