Switching the system off: complete shut down in the main
menu
Select Options > End Software Session in the main menu at the monitor
trolley.
– The System Message dialog box is displayed.
Click Shutdown.
– The imaging system is completely shut down.
Press the OFF switch on the monitor trolley.
– The C-arm system, the imaging system and the monitors are switched off.
Using the End Software Session menu command (or automatically upon
16th switch off using the OFF button), the imaging system is completely shut
down before it switches off. When the ARCADIS Varic is switched on again after
a complete shutdown, it may take more than 3 minutes, depending on the con-
figuration and previous use, for it to reach operational readiness.
The Restart button in the System Message dialog box only shuts down the syn-
go user interface and the application programs and then restarts them. This can
be used, for example, when installing a virus protection update (
Maintenance, Page 10).