Checking The Emergency Stop Function For Motor-Driven System Movements; Maintenance Plan For Checking The System - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Maintenance
Checking the system
Procedures to be performed
Mechanical safety
Electrical safety
Functional check
Image quality
4
of 16
Checking the EMERGENCY STOP function for motor-driven
system movements
Switch the ARCADIS Varic on.
Move the lifting column and press the EMERGENCY STOP button at the
same time.
– The lifting movement is stopped.
– A message is displayed on the monitor indicating that EMERGENCY STOP
has been actuated.
Unlock the EMERGENCY STOP button again.
Press one of the keys for moving the lifting column.
– The lifting movement is enabled again.

Maintenance plan for checking the system

The tests and inspections required by national laws or regulations, such as
DHHS regulations or RöV (constancy tests), are not part of the activities listed in
this maintenance plan.
If national laws or regulations specify more frequent checking and/or mainte-
nance, this must be observed.
Maintenance work should be performed by trained technical personnel only.
To keep the system in an optimum condition, we recommend that you conclude
a maintenance contract. In the case of questions relating to maintenance/mainte-
nance contract please contact our Siemens Customer Service.
Please observe the relevant information in
(
Register 2: Safety, Page 5).
Function
Damage to the housing, system movements and
options (e.g. spacer)
Ground conductor, grounding strap on the C-arm sys-
tem, cables and plugs
Emergency stop and lifting column
Image display and image processing
The stated functions are minimum requirements.
Interval
12 months
12 months
12 months
12 months
Operator Manual
SPR2-310.620.30.02.02