Technical Data
Emissions test
RF emissions
according to CISPR 11
RF emissions
according to CISPR 11
Harmonic emissions
according to IEC 61000-3-2
Voltage fluctuations/
flicker emissions
according to IEC 61000-3-3
12
of 34
Guidelines and manufacturer's declaration –
Electromagnetic emissions
The system is intended for use in an electromagnetic environment as specified
below. The customer or the user of the system should ensure that it is operated
in such an environment.
Compliance
Group 1
Class A
Not applicable
Complies with the speci-
fied regulation
The device or system should not be operated standing next to or on/under other
devices. If adjacent or stacked use is necessary, the equipment or system should
be observed to verify normal operation in the configuration in which it will be
used.
Electromagnetic environment - guidance
The system uses RF energy exclusively for its internal function.
Therefore, its RF emissions are very low and not likely to cause
any interference in nearby electronic equipment.
The system is not suitable for use in domestic establishments
and establishments directly connected to a public power supply
network which supplies buildings used for domestic purposes.
The system is a professionally used unit with a total rated power
larger than 1 kW. There are no limit values for this rated power.