Technical Data
SIREPHOS single-tank
high-frequency generator
Inherent filtration
X-ray tube
Tube voltage
Tube current
Exposure time
Pulsed output
Mean power
18
of 34
X-ray tube unit
Inverter frequency 15 kHz up to 26 kHz
3 mm Al eq. + 0.1 mm Cu
Stationary anode, focal spot nominal value 0.6
Max. output and focuses in operating modes
CFC/SUB/
PFC (> 2 f/s)
ROADMAP/PFC
(≤ 2 f/s)
40 – 110 kV
40 – 110 kV
0.2 – 15.2 mA
1 – 23 mA
n.a.
7 – 24 ms
n.a.
max. 2.3 kW
max. 550 W
max. 550 W
max. 1000 W for
max. 1000 W for
CFC and PFC
PFC high contrast
high contrast
DR (k = 1)
DR (k > 1)
40 – 110 kV
40 – 110 kV
1 – 23 mA
0.2 – 14.3 mA
7 – 42 ms
n.a.
max. 1.8 kW
n.a.
n.a.
n.a.
SPR2-310.620.30.02.02
0
0
Operator Manual