System Description
Preparing the C-arm
Disconnect the power
plug
38
of 44
Transport and parking position of the C-arm system
Prior to transport, the C-arm system must be set to the transport position.
Release all brakes of the C-arm.
Set the C-arm to the transport position shown in the drawing.
– Angulation 0°/Orbital position 0°.
Lower the lifting column down to position 2.
(
Page 14)
Move the horizontal carriage all the way back.
Lock all brakes on the C-arm.
Switch off the ARCADIS Varic and wait for it to shut down.
Pull the power plug out of the wall outlet.
Pull on the plug, not on the cable!
The ARCADIS Varic can be switched on again after approx. 5 s.
Please note that after finishing an examination, the ARCADIS Varic must be shut
down properly before it is disconnected from the power supply.
Operator Manual
SPR2-310.620.30.02.02
0