Product-Specific Notes; Switching On The System; System Power Supply; Printed Circuit Boards - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

6

Product-specific notes

Switching on the system

1. Switch the external main power switch in the room ON.
2. Verify that the generator and BRICK power up properly.
3. Press the power ON button on the control console to activate the MAMMOMAT Nova-
DR
tion
MAMMOMAT Novation
film density display of the control panel to indicate that communication with the MAM-
MOMAT Novation
4. Wait approximately 2 - 5 minutes.
5. Switch the PC and the screen ON at the acquisition workstation.

System power supply

Before you begin working on equipment, it is very important that you disconnect it from
the main power supply at the main circuit breaker. Before removing or inserting any of the
printed circuit boards, switch off the equipment.
WARNING
WARNING

Printed circuit boards

CAUTION
MAMMOMAT Novation DR
. The internal monitoring system automatically performs a functional check of the
DR
. When the detector wing is selected, dr is displayed on the
DR
system is functioning.
If the system is switched off just at the control panel, the line volt-
age is still present at the generator line connection (see wiring di-
agram).
Life-threatening electric shock hazard exists.
Disconnect the mains cable and follow the instructions
presented here.
After the system shuts down, there may still be 380 V DC present
on the intermediate circuit.
Life-threatening electric shock hazard exists.
The printed circuit boards contain highly electrostatically-sensi-
tive components.
The components could be damaged if proper precautions are not
taken.
Use ESD equipment, ground prior to making contact, and
place the components on a conductive surface.
SPB7-250.841.01.04.02
05.05
Prerequisites
Page 6 of 64
CS SD 24
0
0
0
0
Siemens AG
Medical Solutions