Unit Data; Complete System - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Technical Data
General data
Power requirements
Nominal rating
Internal line resistance
Power consumption
Weight
Environmental conditions
for operation
Environmental conditions
for transport
3rd edition IEC 60601-2-28
Classification
Protection against electric
shock
Protection against ingress
of fluids
Operating mode
16
of 34

Unit data

Complete system

100 V, 110 V, 120 V, 127 V, 200 V, 230 V, 240 V ±10%; 50/60 Hz ±1 Hz
20 A to 127 V∼, 15 A from 200 V∼ corresponds to nominal value of the slow-
blow fuse in the power input of the product
Ri < 0.3 ohms at 100 - 127 V∼
Ri < 0.8 ohms at 200 - 240 V∼
max. 2.5 kVA
C-arm system approx. 235 kg
Monitor trolley with 2 monitors, approx. 190 kg (with PC and UPS)
Temperature range: +15 °C to +35 °C
Rel. humidity: 15% to 75%, non-condensing
Barometric pressure: 700 hPa to 1060 hPa
Temperature range: -20 °C to +37 °C
Rel. humidity: 10% to 95%, non-condensing
Barometric pressure: 700 hPa to 1060 hPa
For environmental conditions that fall within the specified values, no wait time is
required for radiation operation.
For environmental conditions that fall outside the specified values, a wait time of
one to twelve hours must be factored in for radiation operation, depending on the
location of the system.
Class 1, no applied part acc. to IEC 60601-1
IPXO (not protected) acc. to IEC 60529
Continuous operation
Operator Manual
SPR2-310.620.30.02.02
0
0