Maintenance
Installing an update in
syngo
10
of 16
As a last step, the update of the virus protection files must be installed and acti-
vated in the syngo user interface.
It is recommended to complete this step immediately to finish the update pro-
cess. For this purpose, the syngo user interface must be restarted. The Software
Distribution/Installation dialog box is automatically displayed to complete the
installation for syngo.
You can also open the Software Distribution/Installation window later (manu-
ally, without restarting). For this purpose, click the yellow exclamation mark on
the status bar. This symbol will appear approx. 15 min after installing the virus up-
date in the system.
Call up Options > End Software Session in the main menu.
– The System Message dialog box is displayed.
Click Restart.
– The syngo user interface and the application programs are shut down and
then started again.
– After the start-up of syngo, the Software Distribution/Installation dialog
box is shown.
Click Install.
– The virus protection update is installed and activated in syngo.
– The Software Distribution/Installation dialog box is closed.
Operator Manual
SPR2-310.620.30.02.02