The Fluoroscopy Tab Card - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Parameter settings
ARCADIS Varic
SPR2-310.620.30.02.02

The Fluoroscopy tab card

The operating programs for continuous fluoro and pulsed fluoro are configured on
the same tab card. There are some parameters that are specific to each of these
operating programs (CONTINUOUS and PULSED areas), and some that are
identical for both (COMMON area).
You can change the following parameter on the tab card of the individual area:
Parameters
Noise Reduction
Noise Reduction Low
Noise Reduction LIH
Remarks
Parameter sets for noise reduction and
or motion detection are possible
Noise Reduction Low is always less or
equal to noise reduction
Special noise reduction factor for
LIH image to guarantee appropriate
image quality even with short-time
release
off (no noise reduction)
k = 2, k = 4, k = 8 (noise reduction on
with selected k factor); with PFC for
pulse rates > 2 p/s only
Configuration
0
15
of 24

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents