Safety Notes For Changing Data - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

External data
ARCADIS Varic
SPR2-310.620.30.02.02
If you change a data set that has also been sent to another network node, please
note the following: The changes affect the local database. The corresponding
data set is not automatically updated on other network nodes. This can lead to
inconsistencies between the local data set and the exported data set.
You have two possibilities to avoid these inconsistencies:
Besides the local data set, you also correct all relevant data sets on other
network nodes.
You delete all relevant data sets on other network nodes and resend the cor-
rected data sets.

Safety notes for changing data

When correcting and rearranging examination data, make sure that the images
are assigned correctly and these changes are also made on the other network
nodes.
Caution
Correcting/rearranging objects with references.
References may be lost!
Rearrange the entire hierarchical group containing all objects with references
in order to maintain the references.
Caution
Rearranging of series/images into another series may lead to wrong image
information, if the selected images/series are not compatible.
Wrong diagnosis due to incorrect image information!
Correct the attributes which do not correspond, before you rearrange the
series/images.
Please terminate the active study of the patient before correcting examination
data of this patient.