ARCADIS Varic
SPR2-310.620.30.02.02
(6) IOD 550 W
SIREMATIC IOD Iodine-contrast optimized
standard mode max. 8.3 mA
in Power Mode 1000 W max. 15.2 mA
SIREMATIC curves
S-1
S-2
LD
HC -1
HC -2
IOD
Fluoroscopy/Pulsed Fluoroscopy
Antiisowatt curve with max. 3 mA
Antiisowatt curve with max. 5 mA
Low dose curve with high kV values: i.e. low radia-
tion exposure for the patient, especially well-suited
for pediatrics.
The characteristically high kV values result in lower
image contrast.
High-contrast curve with max. 4.7 mA
High-contrast curve for applications requiring higher
mA values (max. 7.8 mA) e.g. spinal column, hip,
skull
IODINE curve
For special applications using iodine contrast agent
Technical Data