Radiation Interruption For All Operating Modes - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Safety
22
of 28
I.I. 23 cm/9"
I.I. 23 cm/9"
The values are valid for characteristic HC2 high, 68 kV, 1.3 mA, C-arm horizontal,
downward beam direction without additional Cu filter, with scattered radiation
grid.
I.I. 23 cm/9"
I.I. 23 cm/9"
The values are valid for continuous fluoroscopy 110 kV, 2.9 mA, C-arm horizontal,
downward beam direction without additional Cu filter, with scattered radiation
grid.

Radiation interruption for all operating modes

The hand and footswitches are designed as pushbutton switches. Radiation is
interrupted immediately when releasing the corresponding operating element or,
in other operating modes, after the completion of the stored image.
Height above the floor
[cm]
110
Height above the floor
[cm]
140 to 200
Height above the floor
[cm]
110
Height above the floor
[cm]
140 to 200
Operating area A/
Measurement B
< 1.4 mGy/h
Operating area B/
Measurement D
< 41 µGy/h
Operating area A/
Measurement A
< 9 µGy/h
Operating area B/
Measurement C
< 300 µGy/h
Operator Manual
SPR2-310.620.30.02.02