Cassette Exposure - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Options
Cassette holder
Cassette size
20
of 24

Cassette exposure

The ARCADIS Varic does not have automatic exposure control for cassette tech-
nique.
The cassette holder may only be attached in one specific position at the image
intensifier. In this position the centering pins on the cassette holder (1) must sit
in the recesses in the plastic ring of the image intensifier.
Follow the instructions given on the warning label affixed to the cassette holder!
1
2
Only 24 cm x 30 cm (10 inches x 12 inches).
Caution
The cassette holder is not sufficiently shielded from leaking radiation.
Risk to operating personnel and patient
Do not stand behind the cassette holder in the direction of the radiation.
3
Operator Manual
SPR2-310.620.30.02.02
0