System Description
36
of 44
Reactivating the system
If the imaging system cannot be shut down properly after pressing the OFF but-
ton and does not react any more, you must deactivate the ARCADIS Varic system
as follows:
Open the monitor trolley drawer partway and press the Reset button on the
bottom of the key panel.
– All running processes are aborted and the ARCADIS Varic is shut down.
Switch the ARCADIS Varic on again and let it boot up completely.
– Now you can either continue using the ARCADIS Varic or shut it down.
If the ARCADIS Varic is not fully operational despite the reset, please notify
Customer Service.
Operator Manual
SPR2-310.620.30.02.02
0