System Description
Continuous
Fluoroscopy (CFC)
Pulsed
Fluoroscopy (PFC)
Digital
Radiography (DR)
26
of 44
Press this key.
– Fluoroscopy mode is selected.
– The LED lights up.
Upon startup of the ARCADIS Varic, the basic setting configured for the body
region "All", which is usually Fluoroscopy, is used as the automatic default
(
Register 8: Configuration, Page 5).
Press this key.
– Pulsed Fluoroscopy mode is selected.
– The LED lights up.
Digital Radiography (DR) provides an electronic instant image with best image
quality. DR is recommended for final exposures.
On activation of digital radiography a short radiation pulse is released.
Press this key.
– Digital Radiography mode is selected.
– The LED lights up.
In order to ensure sufficient image quality, images are completely acquired after
the start of acquisition, even if no more radiation is released. The effective acqui-
sition time depends on the preset integration factor and is a maximum of 1.8 sec-
onds.
After complete image acquisition, radiation is automatically switched off, even if
the radiation release button remains pressed.
Operator Manual
SPR2-310.620.30.02.02