The following performance assurance tests check the operation of the CCM. These tests should
be performed upon completion of installation and setup, whenever an assembly of the CCM has
been changed, or as part of troubleshooting to verify proper system operation.
The performance assurance tests check the following functions:
1. Equipment power on and self-test
2. Admit and discharge functions
3. Print and record functions
4. High and low ECG alarms from the bedside
5. If arrhythmia is present:
a. Admit and assign arrhythmia functions
b. Learn signal functions
c. Alarm functions, including alarm validation
d. Paced signal functions
-..,
To check the Admit
Function
1.
2.
3.
4.
5.
6.
7.
h?SS
(Admit/Discharge].
press
~~~~~~~~
Select bed
Enter a name with the keyboard.
Choose a wave.
If you have arrhythmia monito,ring,,,..eroceed to the next procedure. If you do not have
arrhythmia monitoring, press ~~pijii:,,~.~i;t?. The screen should display the patient's
name and the wave selected. Proceed to the procedure to check the recording and print
functions
10-10