Preparing Cable For Termination; Video Cable - HP M2350A Service Manual

Component central monitor
Table of Contents

Advertisement

Preparing the Site
Preparing and Terminating
Cables
Note
Jill
If required, refer to "Lead Colors for System Cables" and Table 5-5 for full
details on the lead colors and signals.
Video Cable
Use MALE CM pin contacts
Use FEMALE (FI socket conlacts
and connector
plug
and connector receptacle
housmgs this side.
housings thls side.
Video Cable @11092A/M1094A)
Use MALE ovl) pin contacts
Use FEMALE LFI socket contacts
and wnnector
dug
and connector receptacle
housings this side.
housings this side.
Video Cable (M1094B)
Preparing Cable for Termination.
Note
3
If using factory crimped contacts, skip to "Terminating Contacts to Connectors
and Faceplates" below.
1. Slip 9.5 cm (3 3/4 ") shrink tubing 0890-0273 over jacket.
2. Strip jacket 5 cm (2 I'). Be careful not to damage the outer shield.
3. Fold back outer shield and drain wire.
4. Remove filler strips, if present.
5. Strip 3.2 cm (1 l/4 ") from jackets of five coaxial wires.
6. Unravel and twist the five coaxial shields together.
7. Solder a wire lead to the twisted coaxial shields. Use wire from excess cable or any 28 or
26
AWG wire.
8. Strip 4 mm (l/8 ") from dielectrics of coaxial wires and twist conductor strands tightly. Be
careful not to stretch the dielectric with the stripper.
9. Use 3.2 cm (1 l/4 ") shrink tubing 0890-0874 to cover inner shield so that the inner and
outer shield do not come in contact. Cut away excess inner shield, if necessary.
5-20