Troubleshooting The Power Supply; Troubleshooting The Dc/Dc Converter - HP M2350A Service Manual

Component central monitor
Table of Contents

Advertisement

Troubleshooting
the System
Troubleshooting
the DC/DC Converter
Troubleshooting
the Power
Supply
If the +5.15V, + 12V and -12V power supplies are present on the Global Bus, three LEDs will
be lit on the Utility CPU. It should be noted that though this is an indication of whether the
power supplies are present it does
not
indicate whether the voltages are within specification.
Troubleshooting
the DC/DC Converter
The input and output signals of the DC/DC Converter may be checked, at the connectors, to
ensure that the signals are present and within specification.
Input Connector (15-pin "D" type)
Signal (+/-%)
Pin Number
+6OV DC input supply (+/-lo%)
6, 7, 8 & 15
+6OV DC loop-through supply (+/-lo%)
13 & 14
OV input ground
1, 2, 3 & 9
OV loop-through ground
10 & 11
No signal
4, 5 & 12
Output Connector (64A,Gpin
DIN)
Signal (+/-%)
Pin Number
+ 12v supply (+ /-10%)
lA,C & 2A,C
-12v supply (+/-10%)
20A,C
+5,15v supply (+/-2%)
23A,C to 29A,C incl.
+ 60V loop-through supply (+/-lo%)
31A,C & 32A,C
Power fail signal
16A,C
System reset signal
17A,C
OV supply ground
3A,C to 9A,C incl.
OV loop-through ground
30A,C
9-20