SECTION IV
PERFORMANCE TESTS
4-1.
INTRODUCTION
4-2. This
section
provides instructions for per¬
formance testing the 8414A. If the serial prefix of
your instrument is different from that listed on the
title page of this manual, there are differences be¬
tween your instrument and the instrument de¬
scribed in this manual. See Paragraphs 1-7 and 1-8.
4-3.
Figure 4-1 is the equipment setup for all
performance tests. The procedures in Paragraphs
4-8 through 4-10 check the 8414A performance
for incoming inspection and periodic evaluation.
The tests can be performed without access to the
instrument interior. The specifications in Table 1-1
are the performance standards. Before starting the
performance tests, allow 30 minutes warm-up time
for the 8414A and Network Analyzer.
4-4.
EQUIPMENT REQUIRED.
4-5.
The test instruments and accessories required
to make the performance tests are listed in Table
4-1. Test instruments other than the ones listed can
be used provided their performance equals or ex¬
ceeds the Critical Specifications listed.
4-6. TEST RECORD.
4-7.
Table 4-2 is a performance test record. This
table may be used during the test to record the test
values
obtained,
and
it
provides
a permanent
record of the test values for use at a later time
during calibration or periodic evaluation.
Table 4-1.
Recommended Test Equipment
Item
Critical Specifications
Recommended HP Model
Oscillator
Frequency Range: 280 (±50) kHz
Output Level: variable from 0 to 3.0 Vrms
Output Impedance: 50 to 600 ohms
200CD, 651A
DC Power Supply
Output: -5 Vdc
6213A
Oscilloscope
Vertical:
Minimum bandwidth: 5 MHz
Minimum sensitivity: 10 mV/cm
Input: dc and ac
Horizontal:
Range: 1 /jsec/cm to 5 psec/cm
180A
with
1801A
and
1821A
Network Analyzer
No Substitute may be used.
8410A
with
8411A
or
8407A
Sweep Oscillator
Frequency Range: any frequency within the
operating range of the Network Analyzer.
8690A with RF unit as
required (see Note 1)
20-dB Attenuator
.
Impedance: 50 ohms nominal
Attenuation: 20 dB ±8 dB
SWR: 1.3 max (280 kHz)
Connector: BNC
8491A with N to BNC
adapters (see Note 2)
*For 8407A Network Analyzers, Model 8601A Sweep Oscillator may be used.
^Oscilloscope
10:1
divider
probe,
such
as
HP
Model
10003A may be used in place
of
20-dB
attenuator.
4-1