Activating the Standby mode (turning off the chamber):
When the function "Standby" is activated, the "Standby" icon is displayed in the screen header in
Normal display. Press the flash icon next to the information icon to display the corresponding text
information "Standby" (information messages, chap. 16.1).
When the controller display is dark in standby mode, press on the touchscreen to activate it for 10 sec-
onds.
Deactivating the Standby mode (turning on the chamber
To deactivate the „Standby" function, unmark the checkbox.
9.7
Controller settings upon start up
Depending on the functions activated in the controller, various settings can be requested directly after
turning on the chamber.
Then you should assign a password for the operating level (chap. 12).
The window „Language selection" enables the language selection, in case that it's activated in the
"Start-up" menu. Afterwards occurs a request of the time zone and the temperature unit.
The controller will function in the operating mode, which was active before the last shut-down. It controls
temperature and pressure in fixed value operating mode to the last entered values and in the program
mode to the set points achieved beforehand.
VDL (E3.1) 06/2020
"Functions on/off" menu
Mark the checkbox of the function "Standby" to activate
it and press the Confirm icon.
In the "Setpoints" menu press again the Confirm icon.
The controller changes to Normal display.
Page 94/206