4.
Completeness of delivery, transportation, storage, and installa-
tion
4.1

Unpacking, and checking equipment and completeness of delivery

After unpacking, please check the chamber and its optional accessories, if any, based on the delivery
receipt for completeness and for transportation damage. Inform the carrier immediately if transportation
damage has occurred.
The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This has
no impact on the function and performance of the chamber.
Please remove any transportation protection devices and adhesives in/on the chamber and on the doors
and remove the operating manuals and accessory equipment.
Due to different installation heights above sea level, a slight negative pressure in the inner chamber may
prevent the door from being opened. Therefore after unpacking the chamber, remove the plug "Manual
ventilation" (7) for emergency ventilation from the tube to ensure pressure equalization. After this, thor-
oughly reinstall the plug.
Risk of injury due to sliding or tilting of the chamber in case of improper lifting.
Injuries, damage to the chamber.
∅ Do NOT lift the chamber using the door or the handle.
∅ Do NOT use technical equipment to lift the chambers.
 Lift chambers size 23 and 56 from the pallet at the four lower corners with the aid of 4
people.
 Lift chambers size 115 from the pallet with the aid of 6 people.
If you need to return the chamber, please use the original packing and observe the guidelines for safe
lifting and transportation (chap. 4.2).
For disposal of the transport packing, see chap. 27.1.
If you ordered the optional pump modulewith a chemical membrane pump, the pump will be delivered in a
separate box and must be fitted into the pump module and connected at the place of installation (see
chap. 6.5.3).
Note on second-hand chambers (Ex-Demo-Units):
Second-hand chambers are chambers that were used for a short time for tests or exhibitions. They are
thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work
flawlessly.
Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before
commissioning the chamber.
VDL (E3.1) 06/2020
CAUTION
Page 50/206