Close All Valves - Binder VDL Series Operating Manual

Vacuum drying oven with microprocessor program controller mb2
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

Press the Setpoint setting icon to access the "Setpoint" setting menu from Normal display..
Setpoint > Fixed value operation setpoints > Functions on/off
Path:
When the function "GAS/AIR 2" is activated, the "GAS/AIR 2" icon is displayed in the screen
header in Normal display. Press the flash icon next to the information icon to display the cor-
responding text information "GAS/AIR 2" (information messages, chap. 16.1)

11.3 Close all valves

Close all existing fine dosing valves
Press the Setpoint setting icon to access the "Setpoint" setting menu from Normal display..
Setpoint > Fixed value operation setpoints > Functions on/off
Path:
When the function "Close all valves" is activated, the "All valves closed" icon is displayed in
the screen header in Normal display. Press the flash icon next to the information icon to dis-
play the corresponding text information "All valves are closed" (information messages, chap.
16.1)
VDL (E3.1) 06/2020
"Functions on/off" menu.
Mark the checkbox of the function "GAS/AIR 2" to acti-
vate it and press the Confirm icon.
In the "Setpoints" menu press again the Confirm icon.
The controller changes to Normal display.
"Functions on/off" menu.
Mark the checkbox of the function "Close all valves" to
activate it and press the Confirm icon.
In the "Setpoints" menu press again the Confirm icon.
The controller changes to Normal display.
Page 105/206

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vdl 23Vdl 23-ulVdl 56Vdl 56-ulVdl 115Vdl 115-ul

Table of Contents