18.7 Value entry for a program section
Main menu > Programs > Time program
Path:
Select the desired program and section.
The section view gives access to all parameters of a program section. You can enter or modify the val-
ues.
The setting and control ranges for the individual parameters are the same as for "Fixed value" operating
mode.
18.7.1 Section duration
VDL (E3.1) 06/2020
Program name and section number
Section duration
Type of setpoint transition: ramp or step
Special controller function
Repeating one or several sections within a program
Temperature setpoint
Temperature tolerance range: minimum and maximum
Pressure setpoint
Pressure tolerance range: minimum and maximum
Section view (partial view).
Select the field "Duration" indicating the time.
"Duration" entry menu.
Enter the desired section duration with the arrow keys
and press the Confirm icon.
Setting range: 0 up to 99 hours 59 min 59 sec.