16.

Notification and alarm functions

16.1 Information messages

Information messages are indicated by information icons displayed in the screen header in Normal dis-
play.
An information icon serves as an indication of a certain condition.
If this condition persists, in some cases an alarm will be triggered after a fix or configurable interval. As
long as the condition persists, the information icon therefore continues to be displayed also in state of
alarm. If during alarm the conditions ends, e.g., if during a tolerance range alarm the actual value returns
to within the tolerance range, the information icon disappears, whereas the alarm will continue until man-
ual acknowledgement.
Press the flash icon next to the information icon to access the corresponding text information.
Information messages overview:
Condition
Deactivated temperature control
Deactivated pressure control
Pressure threshold of 100 mbar not
reached
Ventilation via regular connection
"GAS/AIR" (4)
Ventilation via optional connection
"GAS/AIR 2" (5).
Regular connection "GAS/AIR" (4)
deactivated.
VDL (E3.1) 06/2020
Normal display showing the text infor-
mation.
The currently valid information texts are
highlighted in black (example: "Pressure
threshold not reached" and "GAS/AIR")
Information
Text information
icon
Temperature setpoint
value display shows
" – – – – "
Pressure setpoint value
display shows
" – – – – "
"Pressure threshold not
reached"
"GAS/AIR"
"GAS/AIR 2"
Start after
condition occurred
immediately
immediately
immediately
immediately
immediately
Page 130/206