11.5 Activating / deactivating pressure control
When operating the chamber without a vacuum connection, you can deactivate pressure control with the
to avoid alarms of the pressure system. No pressure tolerance range alarms and no pressure alarm will
be emitted.
Evacuation and ventilation via the controller are then no longer possible. The chamber is ventilated.
Press the Setpoint setting icon to access the "Setpoint" setting menu from Normal display..
Setpoint > Control on/off
Path:
• Checkbox marked: pressure control active
• Checkbox unmarked: pressure control deactivated
The actual pressure value continues to be displayed in Normal display:
VDL (E3.1) 06/2020
"Setpoints" menu.
Select the field "Control on/off" (example: deactivated
pressure control).
Mark / unmark the "Pressure" checkbox to activate /
deactivate pressure control and press the Confirm icon
Normal display with deactivated pressure control.
Page 107/206