Condition
All valves are closed
Chamber is heating up
Standby mode activated
Information messages are not shown in the event list.
16.2 Alarm messages
Alarm messages overview:
Condition
The current actual temperature value
is outside the tolerance range (chap.
15)
The current actual pressure value is
outside the tolerance range (chap. 15)
Safety temperature limiter (TL) has
turned off the heater
Exceeded setpoint of the safety con-
troller class 2
Inner chamber temperature sensor
defective. Heater turns off.
Heater temperature sensor defective.
Heater turns off.
Pressure sensor defective
Failure of temperature sensor for
object temperature (option)
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
VDL (E3.1) 06/2020
Information
Text information
icon
"All valves are closed"
"Heating active"
"Standby"
Alarm message
"Temperature range"
"Pressure range"
"Overtemperature"
"Safety controller"
Actual temperature value
display shows e.g. "- - - - " or
"<-<-< " or ">->->"
"Heater temp.sensor"
Actual pressure value display
shows e.g. "- - - - " or
"<-<-< " or ">->->"
Object pressure value dis-
play shows e.g. "- - - - " or
"<-<-< " or ">->->"
Start after
condition occurred
immediately
immediately
immediately
Start after
condition occurred
after configurable time
(chap. 15.1)
Factory setting: 10 minutes
after configurable time
(chap. 15.1)
Factory setting: 10 minutes
immediately
immediately
immediately
immediately
immediately
immediately