North American Dräger Exclusion System - Dräger Narkomed 2B Operator's Instruction Manual

Anesthesia system
Hide thumbs Also See for Narkomed 2B:
Table of Contents

Advertisement

Section 5 - Operation
Vaporizer
North American
Dräger Exclusion
System
5-2-2
OPEN FUNNEL FILLER
A cam and lever exclusion (interlock) system incorporated into the
vaporizer bank prevents more than one vaporizer from being activated at
a time. The exclusion system requires all unused vaporizers to be locked
in their zero percent positions.
WARNING: Only one vaporizer can be activated at a time. If the
exclusion system permits simultaneous activation of more
than one vaporizer, DO NOT use the anesthesia machine.
Contact a North American Dräger qualified technical
service representative for adjustment.
OP10603
KEY INDEXED
SAFETY SYSTEM