Vaporizers; Vaporizer Exclusion Systems; Dräger Auto Exclusion 2 Vaporizer Mount - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

System Overview

Vaporizers*

The Dräger Vapor anesthetic agent vaporizers are
used to enrich the fresh gas with a precisely mete-
red quantity of vapor from the liquid anesthetic
agent being used, i.e., Isoflurane, Halothane, Enflu-
rane, Sevoflurane, or Desflurane.
When using a Desflurane vaporizer, it must be con-
nected to mains power.
For complete information, consult the appropriate
Instruction for Use provided with the vaporizer.

Vaporizer Exclusion Systems

The exclusion systems available for the Fabius OS
are described below.
Dräger Auto Exclusion 2 Vaporizer Mount*
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When one of the two vaporizers is selected for use
(opened), the interlock mechanism within that
vaporizer's mounting system is activated automati-
cally, preventing the other vaporizer from being
used.
NOTE
Only Dräger vaporizers labeled as
"AUTO EXCLUSION" vaporizers are compatible
with the Dräger Auto Exclusion 2-Vaporizer Mount.
See table for the Auto Exclusion Vaporizer techni-
cal data.
Normal
10 L/min
Operating
Range
Extended
>10 to
Operating
15 L/min
Range
*
optional
20
Dräger Vapor 2000
Instruction for Use
Manual's delivered
concentration accu-
racy values apply.
Dräger auto exclusion
vaporizer concentration
output accuracy may
be reduced.
When using a Desflurane vaporizer, it must be con-
nected to mains power.
Instructions for Use Fabius OS SW 3.n