Electrical Requirements; Electrical Service Specifications - Thermo Scientific Nicolet Summit User Manual

Ftir spectrometers
Hide thumbs Also See for Nicolet Summit:
Table of Contents

Advertisement

Thermo Scientific

Electrical Requirements

Power supplied to the spectrometer must be from dedicated, uninterrupted sources and free
from the following issues:
• Voltage dropouts
• Transient spikes
• Frequency shifts
• Other line disturbances
If you suspect problems with your power, we recommend a power quality audit. Contact us or
your local electrical authority for more information.

Electrical Service Specifications

The following table lists specifications for electrical service. Contact our service representative
in your area if you have questions about the requirements.
Requirements
Input Current
Input Voltage
Line Frequency
Line Disturbances
Noise
Specification
1.6 A Max
100–240 VAC
47–63
Sags, surges, or other line disturbances must not exceed 10% of
input voltage
Less than 2 V (common mode)
Less than 20 V (normal mode)
Before Your Summit Spectrometer Arrives
Nicolet Summit User Guide
6