Thermo Scientific Nicolet Summit User Manual page 61

Ftir spectrometers
Hide thumbs Also See for Nicolet Summit:
Table of Contents

Advertisement

Thermo Scientific
5. Insert the desiccant cartridge.
a. Remove the new O-ring from the packet and install it on the instrument base plate.
Press, rather than roll, the new O-ring into place to be sure it is properly seated in the
groove.
b. Insert the desiccant cartridge into the spectrometer so that the text on the humidity
indicator can be read by someone standing in front of the instrument.
Make sure the cartridge is seated properly over the O-ring and then use the #2
Phillips head screwdriver to tighten the two captive screws.
6. Turn on the spectrometer.
a. Plug in the power cord to the instrument and turn on the power.
b. Reconnect the purge line to the instrument, if applicable, and turn on the purge.
7. Verify the instrument performance.
a. When the instrument is warmed up, open OMNIC Paradigm software and run the
Nicolet FTIR - Factory Qualification workflow.
Replace the Desiccant
Nicolet Summit User Guide
57