d. Snap the male inlet of the wall plumbing assembly firmly into the quick release
fitting.
2. Replace the desiccant cartridge.
a. Turn off the instrument power, unplug the power cord, and remove all accessories
and samples from the system.
WARNING Before opening the desiccant compartment, turn off the instrument
power, unplug the power cord and remove all accessories and samples from the
system. Entry of flammable liquids or gases into the desiccant compartment is an
explosion hazard. If such entry occurs, contact us immediately and do not apply
power to the instrument until the condition has been corrected.
Install and Maintain a Purge Kit
1. To accessory
2. To spectrometer
3. Flowmeter
4. Pressure regulator
5. Quick release fitting
6. Wall plumbing
assembly
Nicolet Summit User Guide
40