a. Use the 0.05-inch hex key to loosen (approximately two turns) the eight screws that
hold the two saturated desiccant canisters to the desiccant cartridge. Remove the
canisters.
b. Turn the desiccant cartridge upside down and peel the old humidity indicator off the
window. Discard the used indicator.
NOTICE Always wear lab gloves or finger cots, or use laboratory tissue, when
handling a new humidity indicator. Oil or moisture from skin can discolor the
indicator.
4. Install the new desiccant canisters.
a. Open the sealed packet that contains the new desiccant (must be sealed to maintain
the desiccant) and remove the new humidity indicator.
b. Turn the desiccant cartridge upside down and press the blue indicator onto the
window so that the flat edges of the indicator are aligned with the edges of the
cartridge. The text should be visible through the window when you turn the cartridge
right side up.
NOTICE Make sure that the indicator is pressed tightly against the window so
that it does not come loose in the desiccant compartment when you reinstall the
cartridge.
c. Insert the canisters into the channel of the desiccant cartridge and tighten the eight
screws (approximately 2 turns) until they are flush with the cartridge channel.
1. Desiccant cartridge
2. Remove these screws (four on
each side)
3. Desiccant canisters
Nicolet Summit User Guide
Replace the Desiccant
56