c. Use a #1 Phillips head screwdriver to fully loosen the three captive screws that hold
the source in place.
d. Swing the bale out. Steady the instrument with one hand and pull firmly on the bale
to pull the source straight out of the instrument.
NOTICE Never touch a source element with your bare fingers. Skin oils or other deposits
on the element will shorten its life. Always use clean finger cots, gloves, or clean laboratory
tissue when handling a source element.
3. Insert the new source.
a. Make sure the three captive screws are straight and then slowly insert the source into
the cavity until the source mount is flush with the instrument.
b. Tighten the screws to secure the source and snap the bale into its storage position.
4. Reconnect and power on the spectrometer.
a. Carefully return the instrument to its upright position and reconnect any cables you
removed.
b. Replace the desiccant. See
Because the internal optics are exposed to the atmosphere when removing the source,
the desiccant should always be replaced after replacing the source.
Replace the Desiccant
for instructions.
Replace the Source
Nicolet Summit User Guide