Align the Spectrometer Optics
Thermo Scientific
Align your spectrometer's optics if a qualification workflow, such as the Nicolet FTIR - PV
test, indicates a failing test. Aligning the spectrometer optimizes the energy that reaches the
detector and maximizes the detector signal. If a performance or qualification test fails, align
the spectrometer's optics, calibrate the laser frequency, and run the test again. If the test still
fails, contact your local technical support representative for help.
To align the spectrometer optics (touchscreen)
1. On the home screen, touch the Diagnostics icon [
2. Open the Align tab and touch Align.
The system automatically aligns the spectrometer and indicates when alignment is
complete.
To align the spectrometer optics (desktop)
1. Select Acquire Data > Diagnostics > Align Spectrometer.
2. In the Align Spectrometer dialog, click Start.
The system automatically aligns the spectrometer and indicates when alignment is
complete.
] to open the diagnostics view.
Nicolet Summit User Guide
50