5. Repeat steps 3 through 4 above to replace the second window.
6. Turn on the instrument power and start OMNIC Paradigm software.
7. Wait at least 15 minutes for the spectrometer to warm up (one to six hours for best
results).
8. Run the performance verification (PV) workflow for the spectrometer.
This workflow, named "Nicolet FTIR-PV Test," is provided with the software.
9. If you use operational qualification (OQ) tests to track instrument performance, we
recommend that you rerun your chosen OQ test after the sample compartment windows
have been replaced (especially if you changed the window type). See the article titled
"Qualify the Summit
10. Replace any accessories you removed earlier on the spectrometer.
Spectrometer" in the online help for more information.
Replace the Sample Compartment Windows
Nicolet Summit User Guide