c. Turn off the purge, if applicable, and disconnect the purge line to the instrument and
any accessory.
Note You can use either desiccant cartridge (with or without the purge
connector) to maintain the required humidity. The illustrations below show the
nonpurged version of the desiccant cartridge. The instructions for replacing the
desiccant canisters are the same for both cartridges.
2. Remove the desiccant cartridge.
a. Remove any installed accessory and all samples from the instrument.
b. Use the #2 Phillips head screwdriver to loosen the two captive screws from the
desiccant cartridge, and lift the cartridge straight up and out of the instrument.
c. Remove and discard the large O-ring.
Note Be careful to not drop the O-ring into the desiccant compartment.
3. Remove the desiccant canisters and replace the humidity indicator.
WARNING The contents of the desiccant canisters (hydrated metal aluminosilicate
molecular sieve) could be harmful if ingested. If you discard the saturated desiccant
canisters, make sure they are properly disposed of.
Replace the Desiccant
Nicolet Summit User Guide
55