Monthly Maintenance; Check The Humidity Indicator; Location Of Humidity Indicator; Humidity Indicator States And Recommendations - Thermo Scientific Nicolet Summit User Manual

Ftir spectrometers
Hide thumbs Also See for Nicolet Summit:
Table of Contents

Advertisement

Monthly Maintenance

Thermo Scientific

Check the Humidity Indicator

The spectrometer's optical components can be easily damaged by excessive moisture in the air.
The spectrometer is sealed and its components are protected by two desiccant canisters that
absorb moisture. A humidity indicator located here monitors the humidity level inside the
spectrometer.
Figure 1.

Location of Humidity Indicator

Check the humidity indicator at least once a month and change the desiccant cannisters when
needed (see the table below), or purchase and install a purge kit. See the article titled
and Maintain a Purge
Kit" for more information.
Table 1. Humidity Indicator States and Recommendations
Humidity Indicator State
Blue
Light blue
Pink or white
For more information, see the article titled "Replace the Desiccant" in the online help.
Meaning
Desiccant is fully charged
Desiccant is becoming saturated
with moisture and no longer
providing sufficient protection
Desiccant has expired
Maintenance Schedule for the Summit Spectrometer
Action
none
Replace the desiccant
Replace the desiccant and
the humidity indicator
Nicolet Summit User Guide
"Install
48