Calibrate the Laser Frequency

Thermo Scientific
Calibrate your laser frequency if a qualification workflow, such as the Nicolet FTIR - PV test,
indicates a failing test. When a test fails, first align the spectrometer, then calibrate the laser
frequency and run the test again. If the test still fails, contact your local technical support
representative for help.

To calibrate the laser frequency using the desktop interface

1. Select Acquire Data > Diagnostics > Laser Calibration.
2. In the Laser Calibration dialog box, click Start to calibrate the laser.
When the system is finished calibrating the laser, a message indicates whether the
frequency was changed and displays the new frequency.

To calibrate the laser using the touchscreen

1. On the home screen, touch the Diagnostics icon [
2. Open the Laser Calibration tab and touch Calibrate to begin.
When laser calibration is complete, a message indicates whether the laser frequency has
been changed and displays the new frequency.
Calibrate the Laser Frequency
]to open the Diagnostics view.
Nicolet Summit User Guide
51