DISPlay:MONitor Subsystem
INPutn:CLOCk[:SOURce]?
Returns the programmed value of the input clock source for Port n.
Parameters
Comments
Example
INPutn:DEBounce:TIME <time> | MINimum | MAXimum | DEFault
Programs the input debounce time for Port n. Ports 0 and 1 share the same
debounce circuitry; Ports 2 and 3 share the same debounce circuitry.
Parameters
Comments
Keysight E1459A/Z2404B User and SCPI Programming Guide
Parameter Name
Parameter Type
INPutn
numeric
– The value returned is an unquoted string of
INPut2:CLOCk[:SOURce]?
Parameter Name
Parameter Type
INPutn
numeric
<time>
numeric (floating pt)
DEFault
MINimum
MAXimum
– The available debounce times are:
18.0 S, 36 S, 72 S, 144 S, 288 S, 576 S, 1.13 mS, 2.26 mS, 4.6 mS,
9.2 mS, 18.4 mS, 36.9 mS, 73.8 mS, 148.0 mS, 294.0 mS, 590.0 mS, 1.18
S, 2.36 S, 4.72 S, 9.43 S, 18.9 S, 37.8 S, 75.0 S, 150.0 S, 300 S, 600 S,
1200 S, 2400 S, 4800 S, 9600 S. Any value sent other than those listed is
rounded to the nearest discrete value.
– Debounce time is rounded to the nearest discrete value. For example 16.0
S to 18.4 S is rounded to 18 S, 18.5 S to 36.4 S is rounded to 36 S,
36.5 S to 72.4 S is rounded to 72 S, etc.
– Ports 0 and 1 use the same debounce time, Ports 2 and 3 use the same
debounce time. For n = 0 or n = 1, both Ports 0 and 1 are set; for n = 2 or n
= 3, both Ports 2 and 3 are set.
Keysight E1459A SCPI Command Reference
Range of Values
0, 1, 2, 3
EXT
INT
or
queries the input clock source for Port
Range of Values
0, 1, 2, 3
18.0 Sec through 9600 Sec
Default 18.0 Sec
Minimum 18.0 Sec
Maximum 9600 Sec
Default
0
.
Default
0
18.0S
ec
51