Configuring for Installation

Configuring for Installation
Before installing the module you should verify that the following jumpers and
switches are set correctly.
Keysight E1459A/Z2404B User and SCPI Programming Guide
– Logical Address dip switch
– Interrupt priority jumper positions
– Input threshold levels
– Reset time of the Watchdog Timer
SHOCK HAZARD. Only qualified, service-trained personnel
who are aware of the hazards involved should install,
configure, or remove the module. Disconnect all power
sources from the mainframe, the terminal module and
installed modules before installing or removing a module.
SHOCK HAZARD. When handling user wiring connected to
the terminal module, consider the highest voltage present
accessible on any terminal.
SHOCK HAZARD. Use wire with an insulation rating greater
than the highest voltage which will be present on the
terminal module. Do not touch any circuit element
connected to the terminal module if any other connector to
the terminal module is energized to more than 30 Vac RMS
or 60 Vdc.
MAXIMUM VOLTAGE. Maximum allowable voltage per channel for
this module is 60 Vdc. Up to 115 Vdc or 115 Vac RMS can be
applied from one channel to another or from any channel to
chassis.
STATIC-SENSITIVE DEVICE. Use anti-static procedures when
removing, configuring, and installing a module. The module is
susceptible to static discharges. Do not install the module
without its metal shield attached.
Installing and Configuring the E1459A
19