Baby Mode Disabled" Alarm; Baby Probe 1 Failure" Or "Baby Probe 2 Failure" Alarm; Bad Membrane Switch" Alarm - GE Giraffe OmniBed Service Manual

Table of Contents

Advertisement

4.3.4 "Baby Mode Disabled" Alarm

Cause or Circumstances
Baby mode has been disabled
in the service screen and
baby mode button is pressed.

4.3.5 "Baby Probe 1 Failure" or "Baby Probe 2 Failure" Alarm

Cause or Circumstances
Probe thermistors show
difference > 0.5ºC.

4.3.6 "Bad Membrane Switch" Alarm

Cause or Circumstances
All
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Failure
None
Failure
Defective patient probe
Defective patient probe jack
wiring
Failure
Defective membrane switch
panel
Action
Press alarm silence button to deactivate the alarm.
If you wish to use the baby mode, re-enable it in service
screen.
Action
Replace patient skin probe.
Plug patient probe into probe jack 1. In service mode,
note the difference between readings from P11 and
P12. Plug the same probe into probe jack 2 and note the
difference between readings from P21 and P22. If they
are greater there is resistance in the probe jack harness.
Replace the harness.
Action
Test switches in service mode. Replace membrane
switch panel.
6600-0343-000 104